Customization: | Available |
---|---|
Type: | Infusion Pump |
Material: | Plastic |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
PARAMETER | SPECIFICATION |
Applicable syringes | 10ml, 20ml, 30ml, 50ml, 60ml of any standard |
Accuracy | +/- 2% with correct calibration |
Rate mode | 50ml(60ml): 0.1ml/h-1800.0ml/h (0.1ml/h step) 20ml(30ml): 0.1ml/h-800.0ml/h (0.1ml/h step) 10ml: 0.1ml/h-400.0ml/h (0.1ml/h step) |
Time Mode | SOL. VOL: 0.1-999.9ml (0.1ml step) TIME: 1-1999min (1min step) |
Body Weight Mode: | Volume: 0.1-999.9ml(0.1ml step) DRUG MASS: 0.1-999.9mg 0.1mg step) DOSE: 0.01-99.99ug/kg/min(0.01ug/kg/min step) WEIGHT: 0.1-300kg(0.1kg the least step) |
Total Volume Range | 0.1-1999.9ml (0.1 ml step) |
Purge Rate | 50ml(60ml): 1800.0ml/h, 20ml(30ml): 800.0ml/h, 10ml: 400.0ml/h |
KVO Rate | 1ml/h |
Occlusion pressure | Three level setting LOW, MEDIUM or HIGH |
Alarms | Near empty, Syringe empty, Volume complete (when limit volume is set), Occlusion, Syringe loose, Operation error, Low battery, device failures. Three levels of alarm volumeLOW, MEDIUM or HIGH |
Bolus Rate | 10ml: 10-300ml/h, 20ml(30ml): 10-600ml/h, 50ml(60ml): 10-1200ml/h |
Bolus Volume | Maximum of 25ml |
Interface | RS-232, Reserved port |
Power source | AC power supply: AC 100 to 240V( +/-10%), 50/60Hz; Built-in Lithium Battery: Rechargeable, 11.1V, 2000mAh. 6 hours of continuous operation. |
Accessories: | AC power line, Operational manual |
Protection Class | Class I, Type CF, IPX4 |
Dimension / Net weight | 300mm*240mm*250mm, 3.6kg |
More products: