Steam Sterilizer Automatic Type Autoclave (AM-280B)
- Shift place type of quick-open cover structure
- The maximum working pressure 0.142MPa
- The highest working temperature 126º C
- Pressure-controlled temperature-adjustable facility 105-126º C
- Sterilizer time-adjustable facility 0-60min
- Double scale numerical indication pressure gauge
- Self inflating type seal
- Fixation pressure device
- Automatic discharge of cold air
- Electric heating
- Operating easy and secrity reliable
1. Normal working conditions of the sterilizer
a) Working power supply: AC220 V±22 V. Frequency: 50 Hz±1Hz
b) Environmental temperature: +5ºC~+40ºC
c) Relative humidity: ≤85%
d) Atmospheric pressure: it reads 0 at an atmospheric pressure from 70kPa to 106 kPa
2. Characteristics of the sterilizer
Type of equipment: portable mode
Type of power supply: single-phase, supplied by power grid
Type of overpressure: Type-II
Type of insulation: Type-I
Input power: see Table 1
Operation mode: uninterrupted service
Class of pollution: Grade-II
Electric protection: basic insulation
4. Interlocking device
The sterilizer is equipped with safety interlocking device on its lid. When it is covered and the steam pressure in the pot is equal to or larger than 0.028Mpa, the interlocking device will lock the lid automatically to keep it closed.
5. Control of temperature, time and water level: the sterilizer is equipped with microcomputer that may display temperature and time in digits and give alarm in case of water shortage or power-off.
Technical Parameters
Applicable power supply: AC 220±10%,50Hz
Consumptive power: 2kW±10%
Working pressure: 0.142MPa
Working temperature: 126ºC
Time setting: 0 to 99 hours
Temperature setting: 50 to 126ºC