Customization: | Available |
---|---|
Classification: | Physiological Functions of Diagnosis and Monitoring Equipment |
Type: | ECG |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Performance
|
|
|
Input mode
|
floating and defibrillation protection
|
|
Frequency response
|
0.05Hz ~ 150Hz(-3dB~+0.4dB)
|
|
CMRR
|
>60dB, >100dB(add filter)
|
|
Time constant
|
≥ 3.2s
|
|
Patient leak current
|
<10µA
|
|
Calibration voltage
|
1mV
|
|
Sensitivity
|
2.5, 5, 10, 20, 40 mm/mV ±5%, standard sensitivity: 10mm/mV±2%
|
|
Noise level
|
<15µVp-p
|
|
Input circuit current
|
≤50nA
|
|
Input impedance
|
≥50MΩ
|
|
Sampling accuracy
|
12-bit
|
|
Safety classification
|
class I, type CF and defibrillation-proof applied part
|
|
Recording mode
|
thermal printing system
|
|
Auto record
|
set record according to auto record format and mode, automatically switching leads, measuring and analyzing.
|
|
Rhythm record
|
set record according to rhythm record format and mode, automatically measuring and analyzing
|
|
Manual record
|
set record according to record format, manually switching leads.
|
|
Measurement parameters
|
HR, P-R interval, P Duration, QRS Duration, T Duration, Q-T interval, Q-Tc, P Axis, QRS Axis, T Axis, R(V5), S(V1), R(V5)+S(V1)
|
|
Paper size
|
80mm(W)×20m(L)
|
|
Sampling frequency
|
1000Hz
|
|
EMG interference filter
|
25Hz/35Hz (-3dB)
|
|
Power frequency filter
|
AC 50Hz/60Hz(-20dB)
|
|
Paper speed
|
Auto: 25, 50mm/s, ±5%
Rhythm: 25, 50mm/s, ±5% |
|
Power supply
|
AC: 100V~240V(50/60Hz)
DC: 7.4V/3500mAh rechargeable lithium battery |
|
Manual
|
5, 6.25, 10, 12.5, 25, 50mm/s, ±5%
|
|
Fuse specification
|
two AC time lag fuse(φ5×20mm), T1.6AL250V
|
|
Waterproof degree
|
IPX0
|
|
Working mode
|
continuous working
|